How to Clean the ProPulse Ear Irrigator

Published: 3rd September 2020

The new ProPulse® Ear Irrigator combines modern design with classic ProPulse® features, enabling more effective ear irrigation techniques and better patient care. The Propulse Ear Irrigation System is intended to facilitate the removal of cerumen and foreign bodies that are not hygroscopic from the auditory meatus. It is also intended for removal of discharge, keratin or debris from the external auditory meatus by irrigation with warm water.

Reasons for using this procedure are to correctly treat otitis external where the meatus is obscured by debris, Improve conduction of sound into the ear, where impacted wax is causing an issue and also to remove a cause of discomfort in the ear. This procedure should ONLY be carried out by a suitably trained clinician.

Cleaning the device

External cleaning of the Propulse Ear Irrigator should be done by hand, wiping with a damp cloth only. Apply liquids to the cloth not the unit. Do not immerse the unit in water. Mild detergents and disinfectants may be used externally. Do not attempt to clean the Propulse QrX™ Tip. Use one Propulse QrX™ Tip per patient treatment and discard to clinical waste after use as this reduces the risk of cross infection between patients.

Cleaning Instructions

The importance of using the correct strength cleaning solution cannot be overstated. A solution that is too strong will damage the device. A solution that is too weak will fail to provide the correct level of cleaning and decontamination. Mirage Health Group recommends the use of Propulse CHLOR-CLEAN Tablets. They are easy and effective to use and provide a measured / specific strength of cleaning solution that is safe and kind to the Propulse internal components. Ensure the unit has been cleaned prior to first use.

  • Place warm tap water into the reservoir up to the horizontal line on the front (G5 & PP17), or the 500ml mark (II, III & NG).
  • Place one Propulse CHLOR-CLEAN tablet into the reservoir and allow it to dissolve completely.
  • Once dissolved run the machine until the cleaning solution leaves the handle. This ensures that the cleaning solution has reached all of the internal components.
  • Leave the solution in place for 10 minutes.
  • After 10 minutes remove the reservoir with the remaining cleaning solution and discard.
  • Fill the reservoir with clean, well run, cold tap water and return to the Propulse.
  • Run the Propulse ensuring that all remaining cleaning solution has been flushed through.
  • Remove reservoir, discard water and dry reservoir thoroughly with a paper towel.
  • Return reservoir to the Propulse – it is now ready for use.

Please Note: Damage caused to your Propulse Ear Irrigator by not using the recommended cleaning agent will invalidate your warranty. Other instruments used with the Propulse should be cleaned according to local guidelines.