How to Clean the Laerdal Suction Unit (LSU)

Published: 18th September 2020

You can Clean the Laerdal Suction Unit easily by following these simple instructions.

Dismantle & Clean the Device

Clean the Serres canister if needed according to manufacturer’s instructions for use. Clean the Serres Vacuum Connector by wiping with a damp cloth or a sponge. Do not autoclave or attempt to disassemble. The Serres Vacuum Connector should be replaced regularly.

Dismantle
Dismantle parts to be decontaminated after each use.
The Float Ball can be snapped out of the lid.
Rinse
Rinse all parts under cold running water a total of 3 times.Immerse in warm water
Clean
Thorough rinsing and cleaning are important steps prior to disinfection.
Immerse parts in hot water (60–70°C) containing a mild detergent.
Thoroughly clean all surfaces, use a brush where possible.
Rinse in warm water and allow to dry.Inspect all parts to be visibly clean and dry.

Disinfection of Reusable Parts

Glutaraldehyde – Room temperature / concentration: 2% 60 minutesRinse all parts in warm water. Allow to dry.
Sodium Hypochlorite (not cleared for use in the US) – Room temperature / concentration: 0.5% 20 minutes.Rinse all parts in warm water. Allow to dry.
Virkon – Room temperature / concentration: 1% 10 minutes.Rinse all parts in warm water. Allow to dry.
Steam autoclaving – Autoclave at max. 121 °C 60 minutes.Allow all parts to cool.

Cleaning – LSU Serres

Parts to be cleaned or replaced regularly:

  • Serres Canister
  • Serres Vacuum Connector
  • The disposable parts must be replaced after each use.

Clean the Serres canister if needed according to manufacturer’s instructions for use. Clean the Serres Vacuum Connector by wiping with a damp cloth or a sponge. Do not autoclave or attempt to disassemble. The Serres Vacuum Connector should be replaced regularly.

Clean the LSU Cabinet

Use a minimum amount of liquid to prevent any electrical shock hazard. Do not immerse the Laerdal Suction Unit or allow it to stand in water or other liquids. This can damage the device, and cause electrical shock resulting in injury to persons.

  • Use a cloth or sponge that is dampened with a mild detergent (hand dishwashing liquid or similar) to clean the external surfaces of the LSU.
  • Use a detergent that is compatible with the materials (See full PDF), and follow the detergent manufacturer instructions.
  • Use a cloth or sponge dampened with water and wipe the surfaces again. Dry the surfaces using a clean cloth or a paper towel.