The Propulse® Ear Irrigator is intended to facilitate the removal of cerumen and foreign bodies that are not hygroscopic from the meatus. And also remove discharge, keratin or debris from the external auditory meatus by irrigation with warm water. Please be aware that this product is sold for medical professional use only. 

How to use the device (Operating Instructions)

These instructions are for general use. When required, refer to the detailed information provided further down in this post.

  • Ensure that the warnings and cautions are observed.
  • Ensure that the patient exhibits no contraindications (previous complications, ear infections, surgery etc.) 
  • Ensure the unit has been cleaned prior to first use, and every day prior to use (See here for the full cleaning guide).
  • The device may only be operated whilst connected to the mains supply using the adaptor and footswitch provided.
  • The reservoir should be removed prior to filling.
  • The water reservoir must be filled to the horizontal line on the front of the reservoir.
  • Monitor temperature regularly to ensure that patient comfort and safety is maintained. Refill as necessary.
  • Fit a new Propulse QrX™ Single Use Tip to the Handle.
  • Adjust the Waterflow Switch (6) to an appropriate level.
  • Turn the On/Off Switch (7) to the position marked “I”.
  • Direct the irrigator tip into the noots tank and switch on the machine for 10-20 seconds in order to circulate the water through the system and eliminate any trapped air or cold water.
  • Ensure the water is warm before presentation to the patient. During treatment you can pause the flow by releasing the footswitch.
  • After treatment empty the reservoir and operate the device to purge any residual water.
  • Remove the Propulse QrX™ Tip and dispose of in accordance with local authority guidelines.
  • Turn off the On/Off switch after use and disconnect from the power supply.
  • Clean the Propulse Ear Irrigator unit every morning prior to use, with a Propulse cleaning tablet.
  • The Propulse Ear Irrigator should only be transported in a Propulse approved carry case toprevent damage or contamination.
  • If any changes in performance occur, turn off the Propulse Ear Irrigator, disconnect from the mains electricity supply and DO NOT use. Please refer the machine to Mirage

Please always follow these operating instructions to ensure you use the device correctly and safely. To maintain optimum performance, service the Propulse Ear Irrigator every 12 months. Any service or repairs carried out by unauthorized agencies/organizations will void any implied or stated warranties provided by the manufacturer.

Detailed Information for Product Use

  • Fittng the Footswitch
    The footswitch is connected to the main unit by a socket in the side of the device. The Propulse® Propulse®Ear Irrigator will NOT operate unless the footswitch is connected.
  • Filling the water reservoir It is recommended that:
    i) The water reservoir is removed from the device for filling and that the lid is always in place when the water container is in-situ on the device.
    ii) The water reservoir should be filled to the horizontal line on the front. This helps to eliminate the risk of spillage.
    iii) Water at 40°C is recommended. Higher temperatures increase the risk of scalding and burns to the patient. Lower temperatures increase the risk of patient discomfort and dizziness
  • Fittng the Propulse QrX™ Tip ®
    The Propulse Ear Irrigator is designed to be used only with QrX™ Single Use Tips. Use one QrX™ Tip per treatment.
    i) Remove Tip from packaging – Tips are non-sterile.
    ii) Push Tip into Handle until a click can be heard.
  • To remove a Propulse QrX™ Tip ®
    i) Retract the QrX™ Locking Collar using a thumb.
    ii) Grasp the used QrX™ Tip between forefinger and thumb and pull gently from the QrX™ Handle.
    iii) Dispose of the used Tip in accordance with local authority guidelines. Do not reuse tips.
    Propulse QrX™ Tips are available to purchase in boxes of 100 individually wrapped (non-sterile) tips from your normal Propulse supplier or from Mirage directly (UK only). Propulse QrX™ Tips are clearly branded with the Propulse logo on the tip and its packaging. Only branded Propulse QrX™ Tips should be used with the Propulse Ear Irrigator.
  • Replacing the Mushroom Valve
    The Mushroom Valve is specifically designed to prevent water flowing out of the reservoir whilst filling. Should a replacement Mushroom Valve be required, please follow the steps below:
    Firstly ensure you are using the correct mushroom valve for this model Propulse Ear Irrigator. This will help prevent possible damage to the water inlet valve.
    i) Remove the reservoir from the Propulse Ear Irrigator.
    ii) Remove the old Mushroom Valve from the reservoir and discard
    iii) Insert a new Mushroom Valve without bending or exerting undue force on the legs of the Mushroom Valve, into the reservoir.
    iiii) Check the condition of the ‘O’ ring on the base of the reservoir and if worn replace with a new unit specific ‘O’ ring.
    ​iiiii) Return the reservoir to the machine
  • ProPulse Power Adapter
    Connect the outlet lead of the Power Adaptor to the Power Adaptor socket marked on the end of the product and to the mains electrical supply. Make sure the cord and Power Adaptor are positioned so that they will not be subjected to damage or stress or present a trip hazard. Only use a Propulse® branded Power Adaptor for product safety. unplug the unit from the power source before attempting to clean it externally.

Purchase the ProPulse Ear Irrigator

HCE-UK offers the ProPulse at an amazing price, accompanied by a full warranty and outstanding customer service. However, it is important to note that this product is intended for use by medical professionals only. Please refer to the options provided below:

Buy now with 100 QrX tips   Buy now with 10 QrX tips
 

Want more info about products & services?

If there something you need help with or maybe need to request a quote, please don't hesitate to give us call.

Trusted UK medical supplier for over 20 years

Established in 2000 – Working alongside GP Surgeries, Hospitals & Organisations Worldwide